1.10 REPLACING THE LITHIUM MODULE
If the following message appears on the screen when the unit is switched on:
This means that the data back-up lithium module is flat (average life 10 years).
It is possible to use unit D100S with a "flat" lithium module, but all parameters
and functions entered by the operator will be lost each time the unit is switched
off.
The module is replaced as follows.
- After obtaining a new lithium module from your Sylvac representative place the unit on a table and remove the 4
retaining screws of the cover.
- Touch a water pipe or other object connected to ground to release any static build-up (the inside of the unit is sen-
sitive to electrostatic discharges).
- Remove yellow cover.
- Remove lithium module with a screwdriver:
- Insert new lithium module, replace cover and replace the 4 cover retaining screws.
- When the unit is switched on again, the memory contents lost message will appear once more.
ERR 3
OUT OF MEMORY
enter one key
eine Taste drücken
presser une touche