13.3 Measuring diameters and centerlines.
The probe constant value should already have been
memorized : (see page 13)
1. Normal mode, if necessary press
key, lights OFF
2. Touch the reference plane (P1)
green light ON
3. Press PRESET key quickly (>1 sec.),
to set the display at zero.
Realise the probe.
4. Change to Diameter/Centerline mode by pressing
key, orange light blink.
5. Place the probe inside the bore to be measured and
touch a point (P2) which is well away from the
reversal point.
6. Move the part or the instrument sideways to the
probe. The reversal point is memorized.
7. Touch a point (P3) well away from reversal point.
8. Move the part or the instrument sideways to the
probe. The diameter is displayed as well
as the symbol.
9. Displaying the centerline can be done in
2 ways :
1. by releasing the probe.
(no possibility to ask again measured diameter)
2. by pressing the key
(Yellow light ON)
The centerline is displayed, as well as the symbol.
31.003
P1
P2
P3
0.000
MM
21.000
M
M
0.000
MM
0.000
MM
90.213
M
M
86.696
MM
30.291
MM
31.003
MM
169.938
MM
169.938
MM